Breaking the D-line, Endless possibilities

Multi-wavelength Abbe Refractometer

Multi-wavelength Abbe Refractometer

DR-M series

PRODUCT LINEUP

Model

DR-M

Option

Digital printer
DP-AD(B)

For DR-M series

DP-AD(B)

Multi-wavelength Abbe Refractometerss [Product details]

Multi-Wavelength Abbe Refractometer DR-M2

Refractive index or Abbe number (νd or νe) can be measured at different wavelengths ranging from 450 to 1,100nm.The DR-M2 digitally displays measurement result of refractive index or Abbe number on the LCD.Measurement can be achieved by matching the boundary line with the intersection point of the cross hairs.This refractometer is connectable with Digital printer DP-AD(B) (Optional).
Specifications
Model DR-M2
Cat.No. 1410
Scale Refractive index
Abbe number
Range Refractive index 1.3278 to 1.7379 (Wavelength 450nm)
Refractive index 1.3000 to 1.7100 (Wavelength 589nm)
Refractive index 1.2912 to 1.7011 (Wavelength 680nm)
Refractive index 1.2743 to 1.6840 (Wavelength 1,100nm)
Minimum Indication Refractive index : 0.0001
Abbe number : 0.1
Accuracy Refractive index : ±0.0002
(With test piece at 500 to 650nm)
Measurement Temp. 5 to 50゚C
(With circulating constant water bath)
Display Panel LCD
Light source Halogen lamp
Output For printer, DP-AD (B) (Optional)
Wavelength Range From 450 to 1100nm (with interference filters)
(For measurement at wavelengths ranging from 681 to 1100nm, the Near Infrared Ray Viewer (Optional is required)
Dimensions & Weight 13×29×31cm, 6.0kg (Refractometer)
15×33×11cm, 3.2kg (Light source unit)

· Lighting glass for film measuerment : RE-6191

· Digital printer DP-AD (B)(Dot matrix printer) : Cat.No.3145

· Interference filter 589(D)nm : RE-3520

· Interference filter 486(F)nm : RE-3521

· Interference filter 656(C)nm : RE-3522

· Interference filter 546(e)nm : RE-3523

· Interference filter 480(F`)nm : RE-3524

· Interference filter 644(C`)nm : RE-3525

· Interference filter (select one wavelength) : RE-3526

· Interference filter 450 to 539nm : RE-3526-1

· Interference filter 540 to 680nm : RE-3526-2

· Interference filter 681 to 799nm : RE-3526-3

· Interference filter 800 to 1100nm : RE-3526-4

· Near-infrared ray viewer : RE-9119

· Eyepiece for polarizing : RE-1146

* Calibration Certificate : Contact an ATAGO representative for details.

· Test piece A with m-naphthalene : RE-1195

· Main & secondary prism : RE-16500

· Illuminating lamp : RE-4170

· Pack of desiccant tablets : RE-8100

*Please inform us of a serial number in the case of an order.

DR-M2

Multi-Wavelength Abbe Refractometer DR-M2

Close

Multi-Wavelength Abbe Refractometer DR-M4

Refractive index or Abbe number (νd or νe) can be measured at different wavelengths ranging from 450 to 1,100nm (For measurement at wavelengths ranging from 681 to 1,100nm, the Optional Near Infrared Ray Viewer (RE-9119) is required.The DR-M4 displays measurement result of refractive index or Abbe number digitally on the LCD.Measurement can be achieved by matching the boundary line with the intersection point of the cross hairs.This refractometer is connectable with Digital printer DP-AD (B)(Optional) .
Specifications
Model DR-M4
Cat.No. 1414
Scale Refractive index
Abbe number
Range Refractive index : 1.5219 to 1.9218 (at 450nm)
Refractive index : 1.4700 to 1.8700 (at 589nm)
Refractive index : 1.4545 to 1.8543 (at 680nm)
Refractive index : 1.4260 to 1.8258 (at 1,100nm)
Minimum Indication Refractive index : 0.0001
Abbe number : 0.1
Accuracy Refractive index : ±0.0002
(With the attached test piece at 500 to 650nm)
Measurement Temp. 5 to 50゚C
Display Panel LCD
Light source Halogen lamp, 12V 50W
Output For printer, DP-AD (B) (Optional)
Wavelength Range From 450 to 1100nm (with interference filters)
(For measurement at wavelengths ranging from 681 to 1100nm, the Near Infrared Ray Viewer (Optional is required)
Dimensions & Weight 13×29×31cm, 6.0kg (Refractometer)
15×33×11cm, 3.2kg (Light source unit)

· Lighting glass for film measuerment : RE-6193

· Digital printer DP-AD (B)(Dot matrix printer) : Cat.No.3145

· Interference filter 589(D)nm : RE-3520

· Interference filter 486(F)nm : RE-3521

· Interference filter 656(C)nm : RE-3522

· Interference filter 546(e)nm : RE-3523

· Interference filter 480(F`)nm : RE-3524

· Interference filter 644(C`)nm : RE-3525

· Interference filter (select one wavelength) : RE-3526

· Interference filter 450 to 539nm : RE-3526-1

· Interference filter 540 to 680nm : RE-3526-2

· Interference filter 681 to 799nm : RE-3526-3

· Interference filter 800 to 1100nm : RE-3526-4

· Near-infrared ray viewer : RE-9119

· Eyepiece for polarizing : RE-1146

* Calibration Certificate : Contact an ATAGO representative for details.

· Test piece C with m-naphthalene : RE-1197

· Main & secondary prism : RE-1640

· Illuminating lamp : RE-4170

· Pack of desiccant tablets : RE-8100

*Please inform us of a serial number in the case of an order.

DR-M4

Multi-Wavelength Abbe Refractometer DR-M4

Close

Multi-Wavelength Abbe Refractometer DR-M2/1550

Refractive index or Abbe number (νd or νe) can be measured at different wavelengths ranging from 450 to 1,550nm. This refractometer digitallys display measurement result of refractive index or Abbe number on the LCD. This measurement can be achieved by matching the boundary line withthe intersection point of the cross hairs This is connectable with Digital printer DP-AD (B) (Optional).
Specifications
Model DR-M2/1550
Cat.No. 1412
Scale Refractive index
Abbe number
Range Refractive index : 1.3278 to 1.7379 (at 450nm)
Refractive index : 1.3000 to 1.7100 (at 589nm)
Refractive index : 1.2912 to 1.7011 (at 680nm)
Refractive index : 1.2743 to 1.6840 (at 1,100nm)
Refractive index : 1.2662 to 1.6759 (at 1,550nm)
Minimum Indication Refractive index : 0.0001
Abbe number : 0.1
Accuracy Refractive index : ±0.0002
(With the test piece at 589nm)
Measurement Temp. 5 to 50°C
(Temperature range available of the circulating water)
Display Panel LCD
Light source Monochromatic light source device
Output For printer, DP-AD (B)(Optional)
Wavelength Range From 450 to 1,550nm
(With interference filters)
Dimensions & Weight 13×29×31cm, 6.0kg (Refractometer)
15×33×11cm, 3.2kg (Power supply unit)

· Lighting glass for film measuerment : RE-6191

· Digital printer DP-AD (B)(Dot matrix printer) : Cat.No.3145

· Interference filter 589(D)nm : RE-16501

· Interference filter 486(F)nm : RE-16502

· Interference filter 656(C)nm : RE-16503

· Interference filter 546(e)nm : RE-16504

· Interference filter 480(F`)nm : RE-16505

· Interference filter 644(C`)nm : RE-16506

· Interference filter (select one wavelength) : RE-16507

· Interference filter 450 to 539nm : RE-16507-1

· Interference filter 540 to 680nm : RE-16507-2

· Interference filter 681 to 799nm : RE-16507-3

· Interference filter 800 to 1100nm : RE-16507-4

· Interference filter 1101 to 1550nm : RE-16507-6

· Near-infrared ray viewer : RE-9119

· Eyepiece for polarizing : RE-1146

* Calibration Certificate : Contact an ATAGO representative for details.

· Test piece A with m-naphthalene : RE-1195

· Main & secondary prism : RE-16500

· Pack of desiccant tablets : RE-8100

*Please inform us of a serial number in the case of an order.

DR-M2/1550

Multi-Wavelength Abbe Refractometer DR-M2/1550

Close

Multi-Wavelength Abbe Refractometer DR-M4/1550

Refractive index or Abbe number (νd or νe) can be measured atdifferent wavelengths ranging from 450 to 1,550nm. This refractometer digitallys display measurement result of refractive index or Abbe number on the LCD. This measurement can be achieved by matching the boundary line withthe intersection point of the cross hairs This is connectable with Digital printer DP-AD(B) (Optional).
Specifications
Model DR-M4/1550
Cat.No. 1415
Scale Refractive index
Abbe number
Range Refractive index : 1.5167 to 1.9166 (at 450nm)
Refractive index : 1.4700 to 1.8700 (at 589nm)
Refractive index : 1.4559 to 1.8557 (at 680nm)
Refractive index : 1.4298 to 1.8296 (at 1,100nm)
Refractive index : 1.4211 to 1.8209 (at 1,550nm)
Minimum Indication Refractive index : 0.0001
Abbe number : 0.1
Accuracy Refractive index : ±0.0002
(With the test piece at 589nm)
Measurement Temp. 5 to 50゚C
Display Panel LCD
Output For printer, DP-AD (B) (Optional)
Wavelength Range From 450 to 1,550nm
(With interference filters)
Dimensions & Weight 13×29×31cm, 6.0kg (Refractometer)
15×33×11cm, 3.2kg (Power supply unit)

· Lighting glass for film measuerment : RE-6193

· Digital printer DP-AD (B)(Dot matrix printer) : Cat.No.3145

· Interference filter 589(D)nm : RE-16501

· Interference filter 486(F)nm : RE-16502

· Interference filter 656(C)nm : RE-16503

· Interference filter 546(e)nm : RE-16504

· Interference filter 480(F`)nm : RE-16505

· Interference filter 644(C`)nm : RE-16506

· Interference filter (select one wavelength) : RE-16507

· Interference filter 450 to 539nm : RE-16507-1

· Interference filter 540 to 680nm : RE-16507-2

· Interference filter 681 to 799nm : RE-16507-3

· Interference filter 800 to 1100nm : RE-16507-4

· Interference filter 1101 to 1550nm : RE-16507-6

· Near-infrared ray viewer : RE-9119

· Eyepiece for polarizing : RE-1146

* Calibration Certificate : Contact an ATAGO representative for details.

· Test piece C with m-naphthalene : RE-1197

· Main & secondary prism : RE-1640

· Pack of desiccant tablets : RE-8100

*Please inform us of a serial number in the case of an order.

DR-M4/1550

Multi-Wavelength Abbe Refractometer DR-M4/1550

Close

Circulating Constant Temperature Bath 60‎-C5

The 60-C5 is an external circulation type constant temperature water tank provided with a thermo-module. The temperature can be set in the range of 10 to 60°C and the unit is perfectly suited for connection with a refractometer.An upgraded version of the 60-C4 model, featuring an increased tank capacity of 1.0L and an improved resolution of 0.1℃ which makes it even easier to use.
Specifications
Model 60‎-C5
Cat.No. 1923
Tank Capacity 1.0 lit
Resolution 0.1℃
Temperature Setting Range 10 to 60°C (water)
Temp.control PID control method using a thermo-module
Dimensions & Weight 20.4×33.6×28.9cm, 9.0kg
Flow Rate 6 lit / min
60‎-C5

Circulating Constant Temperature Bath 60‎-C5

Close